WS-280YDB Horizontal Cylindrical Pressure Steam Sterilizer (microcomputer control) Saturated pressure steam is used to sterilize article, for it will liberate a great amount of heat when condensing. To make the articles (need to be sterilized) hot and wet for a certain time, most resistant bacterial spores will be killed, and then sterilization aim is reached. The method is reliable, economical, and easy to be controlled, so it has been used for more than one hundred years. This unit is ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabric glassed, culture media etc. Technical data code name WS-280YDB1Capacity 280L (Φ0.6m x 1m)2Max.work pressure 0.22MPa3Max.work temperature 1344Heat evenness ≤25Temp.range 401346Timer range 1min99min59sec7Dry time range 1min99min59sec8Power consumption 12KW/AC220V.50Hz9Characteristic Auto water-incooling air releasesterilizetimingair releasedrytimingalarm 10Safety Over pressureover temperature protection deviceauto pressure releaselack water and overheat protectionover current power cut off the door be locked when the pressure inner over0.027MPa,with the function of interlock 11Dimension 85×154×187 (cm)12Trans Dimension 96×161×202 (cm )13Weight Net: 460kg / Gross: 700Kg